Cannon-Ubbelohde Four-bulb Shear Dilution Viscometers
Five-Fold Dilution Type
For measurement of intrinsic viscosities extrapolated to zero shear rates. Provide five-fold range of shear rates. Design gives extra mechanical strength. Negligible kinetic energy correction.
Provided with Cannon’s ISO/IEC 17025 certificate of calibration and instruction sheet.
- Minimum sample volume 20 mL. Require liquid bath depth of 280 mm (11 inches)
- Holder sold separately. Refer to Catalog No. 9726-M70, M73, and M76
Size | Catalog No. | Approx. Constant, cSt/s |
Range cSt* |
Shear Rate at wall reciprocal seconds |
---|---|---|---|---|
25 | 9723-M50 | 0.002 | 0.4 to 2 | 82/1650 |
50 | 9723-M53 | 0.004 | 0.8 to 4 | 45/900 |
75 | 9723-M56 | 0.008 | 1.6 to 8 | 24/480 |
100 | 9723-M59 | 0.015 | 3 to 15 | 15/300 |
150 | 9723-M62 | 0.035 | 7 to 35 | 8/160 |
* Ranges in centistokes represent flow times of 200 to 1000 seconds
Ten-Fold Dilution Type
Similar to 9723-M50 series, but provide ten-fold range of shear rates.
Provided with Cannon’s ISO/IEC 17025 certificate of calibration and instruction sheet.
- Minimum sample volume 20 mL. Require liquid bath depth of 430 mm (17 inches)
- Holder sold separately. Refer to Catalog No. 9726-M70, M73, and M76
Size | Catalog No. | Approx. Constant, cSt/s |
Range cSt* |
Shear Rate at wall reciprocal seconds |
---|---|---|---|---|
25 | 9723-R50 | 0.002 | 0.4 to 2 | 82/3300 |
50 | 9723-R53 | 0.004 | 0.8 to 4 | 45/1800 |
75 | 9723-R56 | 0.008 | 1.6 to 8 | 24/960 |
100 | 9723-R59 | 0.015 | 3 to 15 | 15/600 |
150 | 9723-R62 | 0.035 | 7 to 35 | 8/320 |
* Ranges in centistokes represent flow times of 200 to 1000 seconds.