Instructions for calibration and adjustment of your Thermoprobe Thermometer

ICL is the official representative of ThermoProbe for Florida, Georgia, Puerto Rico and the Caribbean area – for sales, repairs, calibration, and replacement parts.

Recalibration of your ThermoProbe thermometer is necessary :

  • Once each year at least (per API Chapter 7)
  • If you have to change the cable & sensor
  • If the readings of your ThemoProbe, checked against a calibrated thermometer, are not within +/- 0.5 °F

If your ThermoProbe indicates “-1” or “1” or “E-1”, this is indicative of an open circuit in the sensor circuit, and normally means that replacement of the cable and sensor is necessary.

Changing the Sensor & Cable

Physical replacement of the cable and sensor is easily accomplished with normal tools – but take care with the connection of the cable to the main board. It is extremely important that the black (or ground) cable wire is connected to the ground cable connection (on the model TP-2 this is the terminal closest to the face of the instrument) and that the white wire is connected to the other terminal. IF POLARITY IS REVERSED THE THERMOPROBE WILL NOT FUNCTION.

Calibration Procedure

1. Prepare the following equipment

  • A bath for 32 °F or 0 °C (a Thermos bottle with crushed ice works just fine)
  • A constant temperature bath for a high temperature (preferable 195 °F or 90 °C). The actual temperature of the bath is not important, but the temperature must be stabile and should be measured by a calibrated thermometer.
  • A third bath, to check mid-scale, approximately 110 °F or approximately 43 °C. Again, the actual temperature of the bath is not important, but the temperature must be stabile and should be measured by a calibrated thermometer.

2. Prepare the ThermoProbe for calibration – carefully inspect the cable and sensor for indications of physical damage (cuts, cracks in the insulation, etc)

Model TP-2
Open the case by removing the four screws, and look for and locate three (3) potentiometers – one is labeled “zero”, another “span” and the third is labeled “high span”. We are not going to adjust the “high span”.

Model TP-5
Remove the two black screws on the front of the case. Inside are two potentiometers; the lower one is the “zero” and the upper one is the “span”

Model TP-3 and TP-7
Remove the three screws on the face and carefully remove the faceplate. Locate the “zero” and “span” potentiometers.

Model TP-8
On the base of the instrument is a rubber bushing which seals access to the calibration potentiometers. Carefully remove it and locate the “zero” and “span” potentiometers.

Place the sensor in the 32 °F (0 °C) bath and allow it to come to equilibrium (3 minutes is usually sufficient). Adjust the “zero” potentiometer so that the unit reads exactly 32.0 °F (or 0.0 °C) Place the sensor in the high temperature bath, wait 3 minutes, and adjust the “span” potentiometer so that the unit indicates the temperature of the bath (+/- 0.1 °F or 0.1 °C) You will in all likelihood need to repeat this process 2 or 3 times until the unit reads correctly at both the high and low temperatures. When it reads correctly at both temperatures, check it in the middle bath – it must read correctly in the middle bath with precision of +/- 0.3 °F (+/- 0.2 °C) If it does not read well in the middle bath, check the low and high calibrations again. If you cannot adjust it to read well in the middle you probably need to replace the cable and sensor.

If you have problems, call us and with pleasure we’ll help you with the procedure!

INSTRUCCIONES PARA CALIBRAR Y AJUSTAR LAS LECTURAS DE LOS TERMOMETROS DIGITALES DE MARCA THERMOPROBE

ICL es el representante oficial de la compania ThermoProbe para Florida, Georgia, Puerto Rico y el area del Caribe, para ventas, calibracion, reparaciones, y suministro de repuestos.

Es necesario calibrar su ThermoProbe:

1) Cada 12 meses por lo menos (segun API Capitulo 7)
2) Si se tiene que cambiar el sensor y cable
3) Si las lecturas del ThermoProbe no tienen precision de +/- 0.5 °F comparado con un termometro calibrado

Si su ThermoProbe le da un ‘-1’ o ‘E1’, normalmente significa que tienes que cambiarle el cable y sensor.

CAMBIO DEL SENSOR Y CABLE

Es muy facil cambiar el sensor y cable – pero tenga cuidado con la coneccion de las guias al main board. Es sumamente importante que la guia negra este conectada al terminal de tierra (en el TP-2 es el mas cercano al display y la guia blanca al otro terminal. Si estan conectadas al reves el ThermoProbe no funcionara.
PROCEDIMIENTO PARA CALIBRACION

1. Preparar el equipo. Sera necesario tener a mano:

  • Un bano para 32.0 °F (hielo molido en un Thermo funciona muy bien)
  • Un bano para temperatura alta, alrededor de 195 °F. La temperatura de este bano no tiene que ser 195 °F exactamente pero debe de ser una temperatura conocida, estable y precisa
  • Otro bano para verificar el ThermoProbe, cerca de 110 °F, con temperatura conocida y estabile.

2. Prepare el ThermoProbe para calibracion e inspeccione el cable por cortes u otro dano.

Modelo TP-2
Abra el estuche por sacar los cuatro tornillos y busque los potenciometros. Hay tres, uno que tiene el nombre “zero”, uno que dice “span” y otro que dice “high zero”. Nunca debe tocar el que dice “high zero”.

Modelo TP-5
Saque los dos tornillos negros del frente del estuche. Adentro de los agujeros se pueden ver los tornillos de los dos potenciometros, el de mas abajo lo vamos a Ilamar “zero” y el de mas arriba lo vamos a Ilamar “span”.

Modelo TP-3 y TP-7
Saque los tres tornillos de la caratula y busque adentro para encontrar los potenciometros “zero” y “span”.

Model TP-8
En el fondo del instrumento queda un tapon de goma negra, saquelo con cuidado para poder ajustar los potenciometros “zero” y “span”. Luego vuelva a poner el tapon con mucho cuidado.

3. Ajuste
Ponga el sensor en el bano de 32 °F, dejelo puesto aproximadamente por tres minutos para que se equilibre. Prenda el ThermoProbe y lea la numeracion. La numeracion debe decir 32.0, si no, ajuste el tornillo del potenciometro “zero” hasta que lea 32.0 °F.

Ponga el sensor en el bano de 195 °F, dejelo por tres minutos para que se equilibre. Prenda el ThermoProbe y lea la numeracion. Debe mostrar la temperatura del bano, si no, ajuste el tomillo del potenciometro “span” hasta que lea la temperatura del bano.

Se tendra que repetir este proceso 3 o 4 veces hasta que no haya que ajustarlo mas.

Ahora es necesario verificar el ThermoProbe a una temperatura conocida cerca de 110 °F. Ponga el sensor en el bano y espere 3 minutos y vea la numeracion; debe de mostrar la temperatura del bano con precision de +/- 0.3 °F. Si no se logra esta precision, verifique otra vez los 32 y 195 °F, si lee bien en las dos temperaturas pero no lee bien en 110 °F, se tiene problemas con el sensor. Pida un sensor nuevo e instaleselo al ThermoProbe,y calibrelo segun el proceso indicado arriba.

Si tiene problemas con la calibracion de su ThermoProbe, Ilamenos al fono: 561 286 7710 y con mucho gusto les ayudaremos.