Accredited
calibration services:

From -196oC
to 1000oC ASTM thermometers, digital thermometers, PRTs,
thermocouples, thermistors

Specific Gravity
hydrometers. API hydrometers, Baume hydrometers,
Kg/M3 (density) hydrometers, Brix hydrometers, Proof hydrometers

ASTM Class 1 and
OIML E-2 and lesser specification stainless steel and brass laboratory weights.
New calibrated weights and re-calibration of weights you currently have.

Thermo-hygrometers,
chart recorders, humidity sensors, data-loggers

Burets, pipets,
volumetric flasks, graduated cylinders, volumetric glassware

Testing and
certification that an instrument complies with specified accuracy
requirements...







| |
Thermo-Hydrograph
|
 |
NIST Traceable Calibrated thermo-hydrograph
(Temperature & Humidity recorder)
This
recorder is ideal for accurate measurement of temperature and humidity.
Microprocessor based electronic sensors and pen actuators assure an accurate
and robust unit able to withstand shock and vibration. COBEX brand. |
|
Applications:
-
Clean
Rooms
-
Paper
Storage
-
Warehouses
-
Offices
-
Laboratories
-
Hospitals
Features:
-
Microprocessor
Based
-
Rugged / durable construction
-
Battery
back-up available
-
1
or 2 SPDT relay contacts
-
6"
diameter chart
-
110V/220V,
50/60Hz standard
-
Case
may be surface or flush mounted
-
Highly
accurate
-
Fast response time
-
Simple push-button calibration
-
Chart
hold down tabs prevent curling of chart paper
-
Remote electronic humidity and temperature sensors
-
Designed
and manufactured in the USA

|
Calibration Information
Calibrations are performed
utilizing ICL’s Thunder Scientific model 2500 humidity generator. The
NIST-developed two-pressure system of humidity generation on which this unit
operates is an intrinsic standard for humidity (see NCSL
International’s RISP-5 for more details) which relies upon the
NIST traceable calibration of its temperature and pressure sensors to
generate humidity values from 10% to 95%RH at temperatures from 0 to 70°C
with system uncertainties approaching +/- 0.50% RH (k=2).
Skilled, factory trained calibration personnel
subject the client’s instrument to a three (humidity) point test, typically
20, 50 & 80%RH at a specified temperature (normally 23C), with a minimum
equilibration time of two hours at each value, with the sensor(s) exposed to
a moving air stream of controlled humidity and temperature at a velocity of
20 liters per minute. The results of the testing (AS FOUND data) are
evaluated for compliance with required tolerances. If needed, adjustments
are performed to optimize the device’s indications and the testing is
repeated to obtain the AS LEFT data.
A multi-page calibration report is issued which
includes all ISO 9000 and ISO 17025 required information, including
procedure followed, AS FOUND readings, adjustments made, if any, AS LEFT
readings, an uncertainty statement, and all NIST traceability information.
Graphs and charts (if applicable for the device tested) and printout(s) of
test parameters during the test(s) are included as supporting documentation.
Specifications
Humidity Accuracy: +/-
5% (20-80%)
Temperature Accuracy:
+/- 1.0 º C (1.8 º F)
Standard Range:
0 to 100 º (F or C selectable) 0 to 100% Relative Humidity
Chart Rotation Speed: 7 day or 24 hour (please specify)
Ambient Operating Conditions:
0 º to +40 º C, 0-95% RH, Non-condensing Relay Contacts:1 Amp @ 120 VAC
Pen type:
Red and Blue Fiber Tipped (Red = Temp, Blue = Humidity)
Battery Backup:
One 9 Volt Alkaline Battery
Battery Backup Life:
Minimum 24 hours |
| ICL Catalog
Number: |
Description: |
|
C6-STD-24
HOUR |
Range 0-100% RH and
0/100F or C (user selectable), 24-hour chart rotation. Supplied with NIST
traceable, ISO/IEC 17025 certificate of calibration. |
Back to Humidity
home page
|