Accredited calibration services: 

Temperature calibration

From -196oC to 1000oC   ASTM thermometers, digital thermometers, PRTs, thermocouples, thermistors

Hydrometer calibration

Specific Gravity hydrometers. API hydrometers, Baume hydrometers,
Kg/M3 (density) hydrometers, Brix hydrometers, Proof hydrometers

Mass (weight) calibration

ASTM Class 1 and OIML E-2 and lesser specification stainless steel and brass laboratory weights.  New calibrated weights and re-calibration of weights you currently have.

Humidity calibration

Thermo-hygrometers, chart recorders, humidity sensors, data-loggers

Volumetric calibration

Burets, pipets, volumetric flasks, graduated cylinders, volumetric glassware

Verification services

 Testing and certification that an instrument complies with specified accuracy requirements...

Recalibration

Our A2LA Scope

About ICL-Company info

Forms & downloads

FAQs and help topics

Sample reports

Request catalog

MMC International

Portable Electronic Gauging Tapes for Petroleum and Chemical Tanks

FLEXI-DIP Gauging Tapes  (Restricted and Open)

Restricted and Open Gauging Tapes provide a convenient and accurate means for measuring
ullage, product temperature and the oil-water interface. The units are configured to operate 
in the trimode, bimode, interface or temperature mode. Each instrument is designed to 
maximize ease of use. A single penetration of each tank can provide comprehensive data 
about internal tank conditions. When the electronic probe senses hydrocarbon, a steady 
tone is heard; when sensing water, it emits an intermittent tone. Temperature is shown on an 
LED display in the hub of the instrument.
 
 

 

ISO/IEC 17025 REPORT OF CALIBRATION INCLUDED!

 

SPECIFICATIONS
Accuracy
  • of Tape: +/- 1/8" (+/- 3mm), +/- 0.01% of F.S. tape LGT (F.S. = Full Scale).
  • of Temperature Display: +/- 0.5° F (+/- 0.3° C) +/- 1 digit error.
  • of Calibration Range: 32° F (0° C) to 176° F (80° C).
  • of Probe: For dryness to within 3/8 inch (10mm) of bottom.

Application:

  • Open gauging ("O" type) (Safe product vapors only).
  • Restricted gauging ("R" type) via MMC vapor-control valve.

Battery: 9-Volt Eveready # 522 or Mallory MN1604.
Operation:
U-Ullage, I-Interface (oil/water), T-Temperature.
Tape Length

  • Dual-Marked: 50 ft. (15m), 75 ft. (25m), 100 ft. (30m), 120 ft. (37m),165 ft. (50m).
  • Single-Sided: 30m, 100 ft. engineering scale (ft., 1/10 ft., 1/100 ft.)
Approvals: Intrinsically safe--FM, BASEEFA, CSA, SAA, NKK (Japan).
All MMC Flexi-Dip Gauging Tape systems come with a sturdy, molded plastic carrying case 
for convenient transportation and safe storage, and include a spare battery and instruction 
manual.

 

ICL Catalog Number: Tape Length Description
2401TR050FFM-STD       50' Restricted Tri-Mode
2401TR15MFFM-STD      15 Meter Restricted Tri-Mode
2401TR100FFM-STD      100' Restricted Tri-Mode
2401TR30MFFM-STD      30 Meter Restricted Tri-Mode

Optional Calibration

Tape Calibration-MMC

NIST Traceable Certificate of Linearity Calibration in accordance with API Chapter 3.

Replacement Barrels

B2562-35-R-Micro  Replacement barrel – (MB, MMB type vapor valves)
B2562-56-R-F50G Replacement barrel – (MMC “A” C50, F50 type vapor valves)
B2562-279-TS-RG Replacement barrel – (Hermetic C2.SS vapor valves)
B2562-151-TS-RG Replacement barrel – (Hermetic A2.SS vapor valves)