Accredited calibration services: 

Temperature calibration

From -196oC to 1000oC   ASTM thermometers, digital thermometers, PRTs, thermocouples, thermistors

Hydrometer calibration

Specific Gravity hydrometers. API hydrometers, Baume hydrometers,
Kg/M3 (density) hydrometers, Brix hydrometers, Proof hydrometers

Mass (weight) calibration

ASTM Class 1 and OIML E-2 and lesser specification stainless steel and brass laboratory weights.  New calibrated weights and re-calibration of weights you currently have.

Humidity calibration

Thermo-hygrometers, chart recorders, humidity sensors, data-loggers

Volumetric calibration

Burets, pipets, volumetric flasks, graduated cylinders, volumetric glassware

Verification services

 Testing and certification that an instrument complies with specified accuracy requirements...

Recalibration

Our A2LA Scope

About ICL-Company info

Forms & downloads

FAQs and help topics

Sample reports

Request catalog

MMC International

Portable Electronic Gauging Tapes for Petroleum and Chemical Tanks

CLOSED (GAS-TIGHT)

 

 

 

MMC is the world's leading manufacturer of portable instruments for measuring ullage, oil-water interface, temperature, and bottom dryness in shipboard and land-based petroleum and chemical tank systems.  All MMC systems are manufactured in the United States and are certified as intrinsically safe by Factory Mutual, BASEEFA, CSA and SAA.

The Portable Closed Flexi-Dip measuring tape is a battery-operated electronic unit that 
is totally gas-tight. When installed in a vapor-control valve mounted on a storage (cargo) 
tank, it forms a tightly sealed system that prevents toxic vapors from escaping and 
creates a closed gauging system for that tank. The tape is marked in metric or english 
units and is anti-static coated (as is the sensor). In addition, the tape is color marked to 
make tape direction and motion easily seen and to ensure full rewind position. A coiled 
grounding cable with large alligator clip is standard. This closed system is recognized by
 many worldwide regulatory bodies as equivalent to a fixed gauging system as long as 
the tape remains in the valve from the beginning until the end of loading or discharging of the tank.
 
 

ISO/IEC 17025 REPORT OF CALIBRATION INCLUDED!

 

SPECIFICATIONS

Accuracy
  • of Tape Reading: Fluid level +/- 1/8 inch (3mm), +/- 0.01% of F.S. tape LGT (F.S. = Full Scale).
  • of Temperature Display: +/- 0.5° F (+/- 0.3° C), +/- 1 digit error.

Ambient Temperature Limits: -20° F (-7° C) to 120° F (49° C).
Product Temperature Limits: -40° F (-40° C) to 180° F (82° C).
Battery: 9-Volt Eveready # 522 or Mallory MN1604.
Tape Lengths: 50 ft. (15m), 75 ft. (25m), 100 ft. (30M).
Typical Weights: 14 lbs. (6.4 KG).
Pressure Limits: 3 psi working, 7.5 psi maximum.
Approvals: Intrinsically safe--FM, BASEEFA, CSA, SAA, NKK (Japan).

 

ICL Catalog Number: Tape Length Description
2401TC050FFM-STD       50' Closed/Gas Tight Tri-Mode
2401TC15MFFM-STD      15 Meter Closed/Gas Tight Tri-Mode
2401TC100FFM-STD      100' Closed/Gas Tight Tri-Mode
2401TC30MFFM-STD      30 Meter Closed/Gas Tight Tri-Mode

Optional Calibration

Tape Calibration-MMC

NIST Traceable Certificate of Linearity Calibration in accordance with API Chapter 3.

Replacement Barrels

B2562-257  Replacement barrel – (MB, MMB type vapor valves)
A2562-309 Replacement barrel – (MMC “A” C50, F50 type vapor valves)
B2562-267 Replacement barrel – (Hermetic C2.SS vapor valves)
B2562-262 Replacement barrel – (Hermetic A2.SS vapor valves)

More MMC models           

MMC Repair and Recalibration

    

   ICL's main page             Contact us